Laboratory Safety Incidents: Gas Cylinders
Explosions Due to Incompatible Contents (top)
Ruptured Gas Cylinder Destroys
Laboratory Hood (top)
Key Learning Points:
1. Do not refill gas cylinders with highly reactive or
corrosive compounds, including the original compound.
2. Limit the amount of highly reactive, toxic, or flammable
chemicals to the quantity necessary for planned experiments.
Incident Description: A steel lecture bottle located within a hood in a laboratory ruptured with explosive force. The explosion occurred at night and no one was present in the laboratory at the time of the incident. There was, however, significant damage to the lab hood. The gas cylinder contained methyl nitrite (CH3ONO), which had been synthesized and transferred to the cylinder by a postdoctoral researcher who had left the lab approximately 45 minutes prior to the explosion. After the methyl nitrite was transferred to the cylinder it was left at room temperature in the hood. Although methyl nitrite is known to be explosive when heated or exposed to flame, these conditions were not encountered during the synthesis or storage. The cylinder had originally been used to ship iodotrifluoromethane (CF3I) from the gas vendor, and most recently had been used to store nitrosyl chloride (ClNO). (Photo 1)
The Damage: The gas cylinder containing the methyl nitrite ruptured violently, blowing the top cylinder cap off at the weld seam, and twisting the body of the steel cylinder. The cylinder was blown through the front sill of the lab hood, while the blast from the explosion shattered the hood sash, blew off the cabinet doors located underneath the hood, and destroyed the back and top of the hood. The force of the explosion even bent a metal support bracket used to secure utility lines located above the hood. Additionally, the blast produced a crack in an adjacent wall (photo), just outside of the hood opening.
There were very few chemicals located either in the hood or in the cabinet underneath the hood, a fact that surely prevented a fire or significant chemical spill. Although it appears that there was a flash fire associated with the explosion, a sustained fire was probably prevented by the use of a flammable storage cabinet located away from the hood. The only chemical spill was an unknown quantity of ammonium hydroxide from a container that was broken by the blast. There was no free liquid remaining, indicating that the ammonium hydroxide volatilized and was captured by the hood, or surrounding cabinet materials absorbed the liquid. (Photo of damage to the lab hood).
The Cause: Although the cause of the cylinder rupture is not known for sure, several contributing factors have been considered. The major suspicion is potential reaction between methyl nitrite and reactive residues in the cylinder. Although the cylinder was evacuated prior to introduction of the methyl nitrite, the cylinder had previously stored iodotrifluoromethane and nitrosyl chloride, both of which could have potentially caused internal corrosion of the cylinder. Another possibility is that small leaks in the cryogenic transfer system may have allowed liquid oxygen to be deposited in the cylinder. Mechanical failure of the cylinder simply due to the pressure of the methyl nitrite is not thought to be likely since the maximum expected pressure was calculated and found to be within the range commonly stored in these type of cylinders. The deformed condition of the cylinder remains also seem to indicate a more violent explosion than would be expected from a simple mechanical failure.
Lessons Learned and Recommendations: Several lessons learned and recommendations were identified by the Principal Investigator, the Department Safety Committee, and the Safety Office.
1. The gas cylinder had been refilled with potentially corrosive gases.
Although steps were taken to evacuate the used gas cylinder, and high heat or pressure was not present, apparently the cylinder contained residues and reactive surfaces that initiated the chemical reaction that led to the explosion. Never refill gas cylinders with highly reactive or corrosive compounds, including the original compound contained in the cylinder.
2. The amount of methyl nitrite synthesized was far in excess of that required for the planned experiments.
Only a few grams of the methyl nitrite were required for the planned experiments; however, approximately 100 grams were synthesized. Limit the amount of highly reactive, toxic, or flammable chemicals to the quantity necessary for planned experiments, or that will be used within a few months. Avoid the use of these compounds whenever possible.
3. The procedure used to synthesize the methyl nitrite was modified slightly from the original published procedure.
Calcium chloride was used as a drying agent instead of the sodium sulfate used in the published procedure; however, this is not thought to have contributed to the incident. Probably more importantly, the reaction was run in a closed reaction.
4. Implement good laboratory safety procedures and emergency response preplanning.
Several general laboratory safety items were identified as either being related to the actual incident, or the response to the incident. These items included: posting of high hazard areas ("designated areas"), prior approval or review of laboratory experiments, proper chemical storage and availability of flammable storage cabinets, and emergency response procedures and training. These are all basic laboratory safety issues. Implementation of the good laboratory practices contained in the Chemical Hygiene Plan, Biosafety Manual, Radiation Safety Manual, and other laboratory safety references will reduce risks to laboratory personnel.
Leaking Gas Cylinder (top)
Carbon Dioxide Release (top)
Key Learning Points:
1. Use caution if you cannot remove a cylinder cap by
hand.
The Health and Safety Office received an urgent phone call, reporting that a 60 pound
carbon dioxide gas cylinder was venting inside a laboratory. According to lab personnel,
the cylinder cap had not been removed from the time it was delivered by by the campus
vendor. When the employees attempted to use the cylinder, they found that they were unable
to remove the cap by hand. They tried to loosen the cap by inserting a long screwdriver
through the slats of the cap for leverage. In doing so, the main valve was opened and it
began to release its contents. Unfortunately, the cap stayed in place. The lab personnel
tried again, unsuccessfully, to loosen the cap, then left the area and notified the Health
and Safety. What did we learn from this incident? NEVER insert a tool into a cylinder cap in an attempt to open it. If you can't remove
the cap by hand, call the vendor for assistance. If they are unavailable call your health
and safety office. While CO2 is not a flammable or highly toxic gas,
respiratory acidosis can occur from exposure to high concentrations. Death from
asphyxiation could occur if the concentration and duration were sufficient. Regardless of
the chemical hazards of a cylinder's contents, the cylinder, or its cap could become a
missile under the right conditions. NEVER attempt to close a venting cylinder.
Instead, contact the Health and Safety office immediately.
Projectile
Cylinder (top)
The One That Got Away
Cylinder Alterations Lead to Projectile Cylinder (top)
Recently, a liquid nitrogen cylinder exploded at a research facility at X, creating $450,000 worth of damage and rending an entire floor of laboratories inactive for over a month. The explosion was attributed to dangerous alterations that had been made to the cylinder.
Fortunately no one was hurt during the early morning blast. However, the same could not be said of the second floor lab in which the tank was housed. The explosion blew tile off the floor for a five foot radius around the tank. The floor tile turned into quarter-sized pieces of shrapnel that became embedded in the walls and doors of the lab. The force of the explosion also cracked the 6 inch thick concrete laboratory floor which held due to the presence of a supporting beam. Since the floor held, the force of the explosion thrust the nitrogen tank upwards through the lab?s concrete ceiling and into the mechanical room above. There it struck two 3 inch water mains and drove them into the concrete roof of the building, cracking it.
The gas cylinder finally came to rest in the third floor mechanical space leaving behind a 20 inch diameter hole in its wake. The laboratory?s entrance door and wall were blown out into the hallway. The remaining walls of the lab were blown 4 to 8" off of their foundations and all but one of the windows, which was open at the time of the explosion, were propelled into the courtyard. An investigation into the incident revealed the pressure-relief devices of the liquid nitrogen tank had been removed and subsequently sealed with brass plugs. The improper retrofit allowed for the rapid expansion of the nitrogen gas which ultimately blew out the bottom of the tank and allowed the tank to take off like a wayward missile.

Figure 1: Laboratory After the Cylinder Explosion

Figure 2: Exploded Cylinder Compared to an Unaffected Cylinder

Figure 3: Hallway outside the Laboratory Following Cylinder Explosion
Faulty Regulator
Explosion (top)
Faulty Gas Cylinder Regulator
Causes Explosion (top)
A chemist was seriously injured in a
laboratory iwhen opening the valve on a compressed gas cylinder. The regulator exploded
and shattered her jawbone causing unconsciousness. The Canadian Ministry of Labour and the
Technical Safety and Standards Association (TSSA) investigated the accident and found a
fault in the regulator. The regulator was about 10 years old and may have been retrofitted
during that time. The following steps should be taken by all laboratory workers to prevent
similar accidents:
Installation of Regulator:
1. |
Wear appropriate personal protective
equipment such as safety glasses or goggles. Make sure the cylinder is tightly secured
before removing the valve cap. |
2. |
Use only the regulator designed for
the gas being used. If the regulator is not the correct one, the connector will not fit.
Do not force connections. |
3. |
Visually check the regulator for
evidence of damage or contamination and remove any foreign material before attaching the
regulator. |
4. |
If the regulator is dropped, it should
be sent to the supplier for servicing. |
5. |
When attaching the regulator do not
use teflon tape. Use a wrench to tighten the connector securely. |
6. |
Keep the cylinder between you and the
regulator when opening the cylinder valve. Do not stand in front or behind the pressure
gauges when applying pressure to the regulator. |
7. |
Open the regulator's pressure
adjusting valve counterclockwise until it feels free. Slowly open the cylinder valve
(opening the cylinder valve quickly may cause damage to the regulator valve and seals).
When the high pressure gauge indicates maximum pressure, open the cylinder valve fully. |
8. |
Test regulator for leaks. A soapy
water solution can be used for argon, nitrogen, hydrogen or air. For other gases consult
supplier. |
9. |
When not in use store the regulator in
a clean place and do not place anything directly on top of the regulator. |
Maintenance of Regulator:
We recommend that you do an internal check and maintenance on the regulator every six
months and that the regulators are sent to the supplier for servicing every 2 years.
Record all tests and maintenance on a Repair Log Form for Regulators.
Visual Inspection:
|
Carefully examine the entire
regulator. |
|
Check condition of outlet and inlet. |
|
Look for worn threads and damage. |
|
Check to see if gauges are damaged. |
Check for Pressure "Creep":
|
Put regulator onto an appropriate
cylinder with at least 1500 psi. |
|
Make sure pressure adjusting valve is
completely free by turning it counterclockwise. |
|
Slowly open the cylinder valve until
it is fully open. |
|
Set low pressure gauge to
approximately 20 psi, then close off the downstream pressure. Record set pressure. Check
pressure after 30 minutes and record the pressure. |
|
If pressure setting has increased,
remove regulator from service and send out for evaluation. |
Functional Test of Regulator:
|
Close regulator by turning the
pressure-adjusting valve counterclockwise until key is fully released. |
|
Close the cylinder valve and drain the
downstream line. |
|
The low pressure gauge will indicate
zero. Record low pressure gauge reading. The high pressure gauge will read full pressure.
Record initial high pressure. |
|
If the low pressure gauge does not
read zero when all pressure is removed, it may be damaged and must be replaced. |
|
Check high pressure gauge reading
after at least 30 minutes. Record high-pressure reading. Any pressure drop will indicate
leakage. |
|
Release pressure in regulator by
turning pressure adjusting-key clockwise. After venting, fully release pressure-adjusting
key by turning it counterclockwise. |
Revision Date: 6/11/2008
url: http://www2.umdnj.edu/eohssweb/aiha/accidents/cylinder.htm
|